In a July 22 notice, the San Diego-based company warned of pumps that may exhibit a higher rate of speaker failure. During normal use, the insulin pump software monitors current flowing through the speaker during use. Measurements that fall within a pre-determined range indicate a functioning speaker. Meanwhile, measurements falling outside the range indicate a speaker failure.
When the measurements land outside the expected range, the system declares a malfunction, referred to as “Malfunction 16.” If the pump declares this malfunction, insulin delivery will stop and the pump will no longer be operational. Malfunction 16 terminates communication between the pump and continuous glucose monitor (CGM), as well as the t:slim mobile app.
If not addressed, the issue can lead to hyperglycemia, which can result in hospitalization or medical intervention. The company reports 700 adverse events and 59 reported injuries to date, with no reports of death.
Tandem identified that certain speaker versions have a higher rate of Malfunction 16 events due to a wiring issue within the speaker. Users can continue using their pump but with added precautions because Malfunction 16 can occur at any time. They should use the t:slim mobile app with push notifications turned on so the app alerts them if the malfunction occurs, the company said.
Additionally, Tandem intends to release a software update aimed at enhancing the early detection of speaker failure. The update also introduces persistent vibration alarms to help reduce potential safety risk. Tandem plans to notify affected pump users when it makes the update available.